Avance* CS2. Precise care. With a personal touch. Exceptional Design. • Central brake. • Top shelf mounting rails. • Metal work surface.
8 pages

220 KB – 8 Pages

PAGE – 1 ============
GE Healthcare Features •Movable 15” touch screen with CARESCAPE* user interface for the unified CARESTATION* user experience with GE CARESCAPE monitors•ecoFLOW provides information that may help clinicians mitigate the risk of hypoxic mixtures during low and minimum flow•ecoFLOW for visualizing agent consumption and help in mitigating wasteful over delivery of fresh gas flow •User configurable ‘Quick Picks’ for rapid selection of FiO 2and total flow combinationsVentilation Options •With ‘Flow Power Inside’ our ventilator engine can ventilate all patient types from neonates to adults•Sophisticated mechanical ventilation and assisted ventilation options including synchronised PCV-VG with pressure support (SIMV PCV-VG) and minimum rate ventilation (CPAP+PSV) •Vital capacity and Cycling lung ventilation procedures Advanced Breathing System (ABS*) •Compact breathing system specifically designed for low flow to help provide fast gas kinetics for rapid wash-in and wash-out •Continual fresh gas flow with fresh gas flow compensation during mechanical ventilation Avance *CS2Precise care. With a personal touch. Exceptional Design•Central brake •Top shelf mounting rails •Metal work surface •Movable display arm for inbound and outboard positioning •Two or three vaporiser positions •Bi-level work surface illumination

PAGE – 2 ============
Physical SpecificationsDimensionsHeight:139 cmWidth: 77 cmDepth:76 cmWeight:147 kg1Top shelf Weight limit: 34 kg/75 lbWidth: 69.7 cmDepth:44 cmWork surface Height:81.7 cm/32.2 inSize:2640 cm2/409 in2Upper left GCX/DO dovetail GCX length:13.2 cmDO dovetail length: 23.2 cm Upper right DO dovetail Length:34 cmLower right GCX rail Length :41 cmDrawers (internal dimensions) Height:17.5 cm/6.9 inWidth: 33 cm/13 inDepth:26.5 cm/10.4 inAbsorber bag arm (optional)Arm length:39.8 cm/15.7 inBag arm height (adjustable):87 cm/34.3 in 113 cm/44.4 inCastersDiameter: 13 cmBrakes: Central Brake Ventilator Operating Specifications Modes of ventilation – standard VCV (Volume Control) Mode with tidal volume compensation Modes of ventilation – optional Pressure Control Pressure Controlled Ventilation-Volume Guaranteed (PCV-VG) SIMV (Synchronized Intermittent Mandatory Ventilation) (volume and pressure) PSVPro* (Pressure Support with Apnea backup) CPAP+PSV (Pressure support mode) SIMV PCV-VG Ventilator parameter ranges Tidal volume range:5 to 1500ml ( Volume Control, PCV-VG and SIMV volume 20 to 1500ml) (PCV modes 5 to 1500ml)Incremental settings:20 to 50 mL (increments of 1 mL) 50 to 100 mL (increments of 5 mL) 100 to 300 mL (increments of 10 mL) 300 to 1000 mL (increments of 25 mL) 1000 to 1500 mL (increments of 50 mL) Minute volume range:Less than 0.1 to 99.9 L/min Pressure (P inspired ) range:5 to 60 cmH 2O (increments of 1 cmH 2O)5 to 1500 mL volume deliveryPressure (P max) range:12 to 100 cmH 2O (increments of 1 cmH 2O)Pressure (P support ) range:Off, 2 to 40 cmH 2O (increments of 1 cmH 2O)1Excludes vaporisers, airway gas module and patient monitor.

PAGE – 3 ============
Rate:4 to 100 breaths per minute for Volume Control and Pressure Control; 2 to 60 breaths per minute for SIMV, PSVPro and SIMV PCV-VG; 4 to 60 bpm for CPAP+PSV (increments of 1 breath per minute) Inspiratory/ expiratory ratio:2:1 to 1:8 (increments of 0.5) Inspiratory time:0.2 to 5.0 seconds (increments of 0.1 seconds) (SIMV, PSVPro and CPAP PSV) Trigger window:0 to 80% (increments of 5%) Flow trigger: 1 to 10 L/min (increments of 0.5 L/min) 0.2 to 1 L/min (increments of 0.2 L/min) Inspiration termination level:5 to 75% (increments of 5%) Inspiratory Pause range:0-60% Positive End Expiratory Pressure (PEEP) Type: Integrated, electronically controlled Range:OFF, 4 to 30 cmH 2O (increments of 1 cmH 2O)Ventilator performance Pressure range at inlet:240 kPa to 700 kPa (35 psig to 100 psig)Peak gas flow:120 L/min + fresh gas flow Flow valve range:1 to 120 L/min Flow compensation range:150 mL/min to 15 L/min Ventilator Accuracy Delivery/monitoring accuracy Volume delivery:> 210 mL = better than 7% < 60 mL = better than 10 mLPressure delivery:±10% or ±3 cmH 2OPEEP delivery:±1.5 cmH 2OVolume monitoring:> 210 mL = better than 9% < 60 mL = better than 10 mLPressure monitoring:±5% or ±2 cmH 2O3Alarm settingsTidal volume (V TE):Low: OFF, 1 to 1500 mL High: 20 to 1600 mL, OFF Minute volume (V E):Low: OFF, 0.1 to 10 L/min High: 0.5 to 30 L/min, OFFInspired oxygen (FiO 2):Low: 18 to 99% High: 19 to 100%, OFFApnea alarm:Mechanical ventilation ON: < 5 mL breath measured in 30 secondsMechanical ventilation OFF: < 5 mL breath measured in 30 secondsLow airway pressure:4 cmH 2O above PEEP High pressure:12 to 100 cmH 2O (increments of 1 cmH 2O)Sustained airway pressure: Mechanical ventilation ON: Pmax< 30 cmH2O, the sustained limit is 6 cmH2OPmax30 to 60 cmH2O, the sustained limit is 20% of PmaxPmax> 60 cmH2O, the sustained limit is 12 cmH2OPEEP and mechanical ventilation ON Sustained limit increases by PEEP minus 2 cmH2OMechanical ventilation OFF: Pmax2O, the sustained limit is 50% of PmaxPmax> 60 cmH2O, the sustained limit is 30 cmH2OSubatmospheric pressure:Paw < -10 cmH 2OAudio pause countdown clock:120 to 0 seconds Ventilator Components Flow transducerType: Variable orifice flow sensor Dimensions:22 mm OD and 15 mm IDLocation:Inspiratory outlet and expiratory inlet(Optional autoclavable sensor available) Oxygen sensor Type: Optional galvanic fuel cell or paramagnetic with Airway Module option PAGE - 4 ============ Non-disturbing gases:Ethanol, acetone, methane, nitrogen, nitric oxide, carbon monoxide, water vapor: Maximum effect on readings:CO 2< 0.2 vol %; O2, N2O < 2 vol %Carbon dioxide (CO 2)EtCO 2:End-tidal CO 2concentrationFiCO 2:Inspired CO 2concentrationCO2waveform Measurement range:0 to 15% (0 to 15 kPa, 0 to 113 mmHg) Accuracy:±0.2 vol % + 2 % of reading Datex-Ohmeda infrared sensor Adjustable low and high alarm limits for EtCO 2and FiCO 2Respiration rate (RR) Measurement range:4 to 60 breaths/min for E-, M-series modules 4 to 100 breaths/min for CARESCAPE modulesDetection criteria:1% variation in CO 2Adjustable low and high alarm limits for respiration rate; alarm for apneaPatient Oxygen (O 2)FiO2:Inspired O 2concentrationEtO2:End-tidal O2concentrationFiO2-EtO2:Inspired-expired difference O2Measurement Measurement range:0 to 100% Accuracy:±1 vol % +2 % of reading Datex-Ohmeda differential paramagnetic sensor Adjustable low and high alarm limits for FiO2and EtO2; alarm for FiO2< 18%Nitrous Oxide (N 2O)Measurement range:0 to 100% Accuracy:±2 vol % +2 % of reading Anesthetic Agent (AA)Halothane, Isoflurane, Enflurane Measurement range:0 to 6% Accuracy:±0.15 vol% +5% of reading Ventilator screen Display size:15 inchPixel format:1024 x 768Battery backup Backup power: Demonstrated battery time is up to 90 minutes when fully charged. Battery time under extreme conditions is 30 minutes.Battery type:Internal rechargeable sealed lead acidCommunication ports RS-232C compatible serial interface Ethernet Datex-Ohmeda device interface solutions port USB port VGA Output Anesthetic Agent DeliveryDelivery Vaporizers: Tec* 6 Plus, Tec 7 Number of positions:2 (3 as an option) Mounting:Tool-free installation Selectatec* manifold interlocks and isolates vaporizers Airway Modules GeneralM-CAiO, M-CAiOV, M-CAiOVX module software version 3.2 or higher; E-CAiO, ECAiOV, E-CAiOVX Size (WxDxH):75 x 228 x 112 mm/ 3.0 x 9.0 x 4.4 inWeight:1.6 kg/3.5 lb Sampling rate:200 mL/min ±20 mL E-sCAiO, E-sCAiOV Size (WxDxH):38 x 205 x 113 mm/ 1.5 x 8.1 x 4.4 inWeight:0.7 kg/1.5 lb Sampling rate:120 mL/min ±20 mL Automatic compensation for atmospheric pressure variation (500 to 800 mmHg for E-, M-series modules; 495 to 795 mmHg for CARESCAPE modules) temperature and CO 2/N2Oand CO 2/O2collision broadening effect. Parameter display update interval typically breath-by-breath. Functional alarms for blocked sample line, D-fend check and D-fend replacement. 4 PAGE - 5 ============ Sevoflurane Measurement range:0 to 8% Accuracy:± 0. vol% + % of reading Desflurane Measurement range:0 to 20% Accuracy:±0. vol% + % of reading Waveform displayed MAC value displayed Identification threshold:0.15 vol % 1Agent mixture detection Adjustable high and low alarm limits for EtAA, FiAA Patient Spirometry* Pressure-volume loop Pressure-flow loop Flow-volume loop Airway pressure and flow waveforms Adjustable low and high alarm limits for Ppeak, PEEPtotandMVexpAlarms for MVexp<< MVinspand for MVexplow. Detection through D-lite* or Pedi-lite* flow sensor and gas sampler with following specifications:Compact Airway Modules D-litePedi-lite Respiration rate:4 to 354 to 50 breaths/minbreaths/min Tidal volumeMeasurement range:150 to15 to 2000 mL300 mL Accuracy1:±6% or 30 mL±6% or 4 mL Minute volume Measurement range:2 to 20 L/min0.5 to 5 L/min Accuracy1:±6%±6% Airway pressure Measurement range:-20 to +100 cmH2OAccuracy1:±1 cmH2ODisplay units:cmH 2O, mmHg, kPa, mbar, hPa FlowMeasurement range:1.5 to 0.25 to 25 L/min 100 L/min1Typical value 5I:EMeasurement range:1:4.5 to 2:1 ComplianceMeasurement range:4 to 1001 to 100 mL/cmH2OmL/cmH 2OAirway resistance Measurement range:0 to 40 cmH 2O/L/sCARESCAPE Airway Modules D-lite(+)Pedi-lite(-) Respiration rate:4 to 354 to 70 breaths/minbreaths/min Tidal volumeMeasurement range:150 to5 to 2000 mL300 mL Accuracy1:±6% or 30 mL±6% or 4 mL Minute volume Measurement range:2 to 20 L/min0.1 to 5 L/min Accuracy1:±6%±6% Airway pressure Measurement range:-20 to +100 cmH2OAccuracy1:±1 cmH2ODisplay units:cmH 2O, mmHg, kPa, mbar, hPa FlowMeasurement range:-100 to -25 to 25 L/min 100 L/minI:EMeasurement range:1:4.5 to 2:1 ComplianceMeasurement range:4 to 1001 to 100 mL/cmH2OmL/cmH 2OAirway resistance Measurement range:0 to 200 cmH 2O/L/sSensor specificationsD-lite/Pedi-lite/ D-lite(+)Pedi-lite(+) Dead space:9.5 mL2.5 mL Resistance at 30 L/min:0.5 cmH2Oat 10 L/min:1.0 cmH 2O PAGE - 6 ============ 6Electrical SpecificationsCurrent leakage 100/120 V: < 300µA220/240 V: < 500µAPower Power input: 100-120 Vac, 50/60 Hz 220-240 Vac, 50/60 Hz Power cord: Length:5 m/16.4 ft Rating:10A @ 250 Vac or 15A @ 125 Vac Inlet/outlet modules 100 -120 V System circuit breakers: 15A Outlets: 4 outlets on back, 3-2A, 1-3A individual breakers, isolation transformer220-240 V System circuit break ers: 8A Outlets (optional): 4 outlets on back, 3-1A, 1-2A individual breakers, isolation transformerPneumatic SpecificationsAuxiliary common gas outlet (optional)Connector: ISO 22 mm OD and 15 mm IDGas supplyPipeline input range:280 kPa to 600 kPa (41 psig to 87 psig)Pipeline connections:DISS-male, DISS-female, AS4059, BSPP 3/8, S90-116, or NIST All fittings available for O 2, N2O, and Air, and contain pipeline filter and check valve Cylinder input:Pin indexed in accordance with CGA-V-1 or DIN (nut and gland); contains input filter and check valve. Note: Maximum 3 cylindersPrimary regulator diaphragm minimum burst pressure:2758 kPa/400 psig Primary regulator nominal output:a/50 psig Pin indexed cylinder connections a/60 psig DIN cylinder connectionsO2controls Method:N2O shut off with loss of O2pressure Supply failure alarm:< 252 kPa (36.55 psig) O2flush:Range: > 5 L/minAlternate O2(safety flow)Range:500 mL/min minimum to 10 L/minIndicator:Flow tube Indicator accuracy:±5% full scale Fresh gas Flow range:0 and 150 mL/min to 15 L/min (minimal flow capable)Total flow accuracy:±10% or ±40 mL/min of setting (larger of) O2flow accuracy:±5% or ±20 mL/min of setting (larger of) Balance gas flow accuracy:±5% or ±20 mL/min of setting (larger of) Air/N 2OO2concentration range:21% to 100% when Air is available O2concentration accuracy:2.5% (T 5% (Total Flow < 1 L/min) 6.5% (Total Flow < 0.4 L/min) Electronic mixer response time:500mS (10% to 90% flow step) Compensation:Temperature and atmospheric pressure compensated to standard conditions of 20°C and 101.3 kPa Hypoxic guard:Electronic MaterialsAll materials in contact with patient breathing gases are not made from natural rubber latex. PAGE - 8 ============ GE Healthcare P.O. Box 900, FIN-00031 GE, Finland GE Direct United Kingdom: +44 (0)800 0329201 www.gehealthcare.com © 2013 General Electric Company – All rights reserved. General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE representative for the most current information. GE and the GE Monogram are trademarks of the General Electric Company.*Avance, ABS, Tec, CARESCAPE, Carestation, D-lite, Pedi-lite, PSVPro, Selectatec and Patient Spirometry are trademarks of General Electric Company.Not for sale in all markets. Please check with your sales representative. Always refer to the complete instructions manuals before use. GE Healthcare, a division of General Electric Company. DOC1183602 Rev 7 /13Healthcare Re-imagined GE is dedicated to helping you transform healthcare delivery by driving critical breakthroughs in biology and technology. Our expertise in medical imaging and information technologies, medical diagnostics, patient monitoring systems, drug discovery, and biopharmaceutical manufacturing technologies is enabling healthcare professionals around the world discover new ways to predict, diagnose and treat disease earlier. We call this model of care “Early Health.” The goal: to help clinicians detect disease earlier, access more information and intervene earlier with more targeted treatments, so they can help their patients live their lives to the fullest. Re-think, Re-discover, Re-invent, Re-imagine. 220 KB – 8 Pages